Ubicación: Global + English
Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Suspended particle testing measures the concentration of non-viable airborne particles in cleanrooms and controlled environments to verify compliance with ISO 14644-1, GB/T 25915.1, and GMP cleanliness classifications. This technical guide explains the measurement principles, applicable standards, sampling methodology, and industry applications of airborne particle count analysis for pharmaceutical, electronics, medical device, and other contamination-sensitive facilities.

Technical Overview

Suspended particle testing, also referred to as airborne particle count testing or non-viable particle monitoring, is the quantitative measurement of solid and liquid particles suspended in the air of controlled environments. Particles of concern typically range in size from 0.1 micrometers (μm) up to 5 μm and larger, since particles within this range can contaminate products, damage sensitive surfaces, and serve as carriers for microorganisms in sterile manufacturing.

Airborne particles originate from multiple sources, including personnel skin and clothing shedding, equipment friction and wear, process-generated debris, packaging materials, and infiltration of outdoor air. Because these particles are invisible to the naked eye at the concentrations relevant to cleanroom operation, dedicated instrumentation is required to detect, size, and count them accurately.

The measurement is most commonly performed using a light scattering airborne particle counter (LSAPC). As air is drawn through the instrument, individual particles pass through a laser beam and scatter light onto a photodetector. The intensity of each scattered-light pulse is proportional to particle size, allowing the instrument to classify particles into defined size channels and report concentration as particles per cubic meter of air sampled. Cleanliness is then verified against the classification limits defined in ISO 14644-1 (identical to the Chinese national standard GB/T 25915.1), which defines nine cleanliness classes from ISO Class 1 (cleanest) to ISO Class 9.

Suspended particle testing is a core requirement of contamination control programs in regulated industries. ISO 14644-2 mandates periodic re-verification of cleanliness classification, while EU GMP Annex 1, WHO GMP guidance, and PIC/S requirements embed particle limits directly into the Grade A–D classification system used for sterile medicinal product manufacturing.

Testing Samples

Cleanroom Air

Air within classified cleanroom spaces is the primary testing target. Testing can be performed in the as-built state (facility complete, no equipment or personnel), the at-rest state (equipment installed and operating, no personnel present), or the operational state (normal production activity with personnel present). The applicable occupancy state must be agreed upon before testing, since classification results differ significantly between conditions.

Clean Zones and Localized Enclosures

This category includes laminar flow cabinets, unidirectional airflow modules, biological safety cabinets, transfer hatches, and restricted access barrier systems (RABS). These smaller volumes require careful probe placement to evaluate the air quality at the critical working plane where product is exposed.

Isolators and Sterile Processing Enclosures

Sealed isolators used for aseptic filling, sterility testing, and handling of hazardous materials are tested to confirm they maintain the tightest particle limits, typically corresponding to ISO Class 5 or better within the critical zone.

Supply Air and HVAC Verification

Terminal filtered supply air, air from fan filter units, and air downstream of HEPA/ULPA filters can be tested to confirm filtration performance and to localize contamination sources when room-level results deviate from specification.

Testing Items

Suspended particle testing programs typically include the following measurable parameters:

  • Particle concentration at specified threshold sizes, most commonly 0.1 μm, 0.2 μm, 0.3 μm, 0.5 μm, 1.0 μm, 3.0 μm, and 5.0 μm, reported as particles per cubic meter
  • ISO Class 1 through ISO Class 9 cleanliness classification verification in accordance with ISO 14644-1 / GB/T 25915.1
  • GMP Grade A, B, C, and D verification at 0.5 μm and 5.0 μm thresholds in accordance with EU GMP Annex 1 and national GMP annexes
  • U descriptor measurement for ultrafine particles smaller than 0.1 μm, requiring condensation nucleus counting instrumentation
  • M descriptor assessment for macroparticles larger than 5 μm, evaluated using dedicated large-particle measurement approaches
  • Time-based trend analysis during operational monitoring to identify contamination excursions linked to processes, personnel, or equipment events

Testing Methods

Instrumentation

Testing is performed with light scattering airborne particle counters calibrated and verified in accordance with ISO 21501-4, which defines requirements for size calibration, counting efficiency, flow rate accuracy, and false count rate. Instruments with sample flow rates of 28.3 L/min (1 CFM) or higher are commonly used for cleanroom classification, while high-flow instruments are preferred in ISO Class 5 and cleaner environments where long sample times would otherwise be required. Condensation particle counters are used when sub-0.1 μm measurement is required.

Determination of Sampling Locations

Per ISO 14644-1:2015, the minimum number of sampling locations is calculated as the square root of the cleanroom or clean zone area in square meters, rounded up to the next whole number. Locations are distributed evenly across the area and positioned at working height, typically 0.8 to 1.5 meters above the floor, unless process-defined heights apply.

Determination of Sample Volume and Duration

The minimum sample volume per location is derived from the formula Vs = (20 / Cn,m) × 1000, where Cn,m is the class limit for the largest considered particle size, expressed in particles per cubic meter. The resulting volume must be at least 2 liters, and the sampling duration must be at least 1 minute per sample. At least three sample volumes are taken at each location, and results are averaged.

Sampling Procedure

An isokinetic sampling probe is oriented directly into the airflow in unidirectional flow zones. In non-unidirectional areas, the probe is oriented upward and positioned away from obstructions, filters, and equipment surfaces that could distort results. A pre-sampling purge is performed between locations, and the counter zero count is verified using a filtered zero-count adapter before and after the measurement campaign.

Data Evaluation and Reporting

Under ISO 14644-1:2015, each sampling location must individually meet the classification limit for the stated particle size thresholds, replacing the 95% upper confidence limit approach used in the 1999 edition. The test report documents the classification result, occupancy state, considered particle sizes, instrument details, calibration traceability, sampling locations, and environmental conditions such as temperature, humidity, and differential pressure observed during testing.

Reference Standards

  • ISO 14644-1 / GB/T 25915.1 — Classification of air cleanliness by particle concentration
  • ISO 14644-2 / GB/T 25915.2 — Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration
  • ISO 21501-4 — Determination of particle size distribution: calibration of light scattering airborne particle counters
  • GB 50591 — Code for construction and acceptance of cleanroom and associated controlled environment
  • EU GMP Annex 1 — Manufacture of sterile medicinal products

Applications

Suspended particle testing supports contamination control, regulatory compliance, and product quality across a broad range of industries:

  • Pharmaceutical and biopharmaceutical manufacturing: classification and periodic requalification of aseptic filling rooms, grading of Grade A–D areas, qualification of cleanrooms prior to production release, and investigation of contamination events
  • Medical device manufacturing: verification of controlled environments for sterile device assembly, primary packaging, and terminally sterilized products
  • Semiconductors and electronics: qualification of ISO Class 1–5 fabrication areas, clean assembly rooms for precision optics, display production, and lithium battery electrode and cell assembly environments
  • Hospitals and clinical pharmacies: verification of preparation rooms, compounding isolators, and operating theater air cleanliness in support of infection control programs
  • Food, cosmetics, and consumer products: environmental control of filling lines, low-bioburden processing zones, and preservative-free product manufacturing areas
  • Aerospace, automotive, and precision machining: particle control in the assembly of fuel systems, hydraulic components, sensors, and optical instruments
  • Cleanroom commissioning and acceptance: performance verification of newly constructed or modified facilities in combination with airflow, filter integrity, pressure cascade, and recovery testing

FAQ

What is the difference between suspended particle testing and microbial environmental monitoring?

Suspended particle testing counts all non-viable inert particles in the air using optical instrumentation, while microbial monitoring collects and cultivs airborne viable organisms using settle plates, active air samplers, or contact plates. The two methods are complementary, because inert particles often serve as carriers for microorganisms. Regulatory expectations typically require both programs to run in parallel.

How often must suspended particle testing be performed?

ISO 14644-2:2015 specifies maximum time intervals for cleanliness re-verification: 6 months for cleanrooms of ISO Class 5 and cleaner, and 12 months for ISO Class 6 through ISO Class 8 rooms. GMP-regulated facilities and site quality systems may impose more frequent schedules, and retesting is always required after modifications, filter replacement, or maintenance that could affect air cleanliness.

Which particle sizes should be tested?

The considered particle sizes depend on the classification target and applicable regulation. ISO 14644-1 classifications are defined for one or more threshold sizes between 0.1 μm and 5 μm, and the sizes selected must be stated in the classification requirement. GMP Grade testing is performed at 0.5 μm and 5.0 μm, with 0.5 μm serving as the primary classification threshold.

Why is ISO 21501-4 calibration important for particle counters?

ISO 21501-4 calibration establishes traceable size response, counting efficiency, and flow accuracy, which directly determine whether measured concentrations can be compared against classification limits with confidence. Regulatory inspectors increasingly expect documented ISO 21501-4 calibration as a condition for accepting particle count data used in qualification and release decisions.

Should testing be performed in the static or operational state?

The occupancy state must match the intended claim. Static (as-built or at-rest) testing demonstrates the capability of the facility and its air handling systems, while operational testing demonstrates real-world performance under production load and reflects the risk to exposed product. Many quality agreements require both states during initial qualification, followed by routine operational monitoring.

Can suspended particle testing be performed by external laboratories?

Yes. Accredited third-party laboratories routinely perform suspended particle testing using calibrated instrumentation and qualified personnel, issuing results in formats suitable for regulatory submissions, customer audits, and internal quality records, provided the laboratory operates under a documented quality management system with traceable calibration.